•Two-stage Cascade Variable Frequency Systems
•UL/CE Certified
•HC refrigerant
•Power Consumption as 8.7kw/24h.
•Temperature Uniformity:3℃
•10-inch Intelligent Touch Screen
•6 sides VIP vacuum Insulation Material
•Two-stage Cascade Variable Frequency Systems
•UL/CE Certified
•HC refrigerant
•Power Consumption as 8.7kw/24h.
•Temperature Uniformity:3℃
•10-inch Intelligent Touch Screen
•6 sides VIP vacuum Insulation Material
M-Cascade system ULT freezer that consists of two stages cascade cooling technology with absolute hydrocarbon refrigerant, which are R290 and R170 as natural gas. The inverter compressors programmed to help energy saving under long period preservation or provide fast temperature recovery while frequently door open to maintain specimens liveness.
Two-stage Cascade & Variable Frequency Drive Refrigeration
High effiency frequency conversion compressors enable the operating power to be automatically adjusted according to the heat load,making the equipment energy saving.
Applicable to Various Power Supply Environments
It can work normally under various power supply environments such as 110V60Hz, 220V50Hz, 220V60Hz, ect.
Stable Internal Temperature
Standard SECOP compressors maintain a stable -86℃ ultra low temperature environment, that increases efficient sample storage.
Excellent Thermal Insulation Performance
The design of foam inner doors, built-in vacuum insulation board and triple independent silicone sealing strip design make the insulation performance excellent.
Human-oriented
· Ergonomic handle, smooth track and high recision pressure relief valve make it easier to open and close the door;
· 3 division plates are equipped inside to divide the storage area into 4 layers for convenient sample placement; 2 internal doors are also equipped to reducing frosting caused by door opening;
· Stainless steel (SUS304) as the internal material, is durable and easy-to -clean;
· Equipped with 4 heavy casters and 2 balancing feet for easy movement and fixation of the equipment.
Two-stage cascade system
±4 ℃ Temp.uniformity
High-definition and easy-handling control panel to set the temperature range from -40°C to -86°C with digital temperature display precision in 1°C
Lockable door, keyboard lock and password access prevents any unauthorized access and operation
Adjustable 3 stainless steel shelves with inner door and double silicon gasket seal
Optional: Chart recorder, CO2 back up system, freezer racks/boxes/tubes, Remote alarm system,Modbus protocol for data transfer, Card access or finger print access,Conform to 21 CFR 11 regulation
China's leading manufacturer of ultra-low temperature freezer
Ultra low deep freezer DW-HL528S can be used as laboratory freezer, ultra low medical freezer and deep freezer,suitable for use in hospitals, health and disease prevention systems, research institutions, colleges & universities, the electronic industry, biological engineering, laboratories in colleges & universities, military enterprises, deep-sea fishing companies, etc.
-86℃ Ultra-low Temperature Freezer | |
Model | DW-HL528S |
Capacity(L) | 528 |
Internal size(W*D*H)mm/Inch | 585*696*1286/ 23.0*27.4*49.8 |
External size(W*D*H)mm/Inch |
930*1041*1947/ 36.6*41.0*76.7 |
Package size(W*D*H)mm/Inch | 1035*1165*2158/ 40.7*45.9*85.0 |
NW/GW(Kgs) | 260/350(wood packing) |
Performance | |
Temperature range | -40~-86℃ |
Ambient temperature | 16-32℃ |
Cooling performance | -86℃ |
Climate class | N |
Controller | Microprocessor |
Display | 10‘’HD Intelligent touch screen |
Refrigeration | |
Compressor | 2pcs |
Cooling method | Direct cooling |
Defrost mode | Manual |
Refrigerant | High stage: R290, Low stage: R170 |
Insulation thickness(mm) | L/R:130,B:130 |
Construction | |
External material |
Cold rolled steel |
Inner material | Stainless steel |
Shelves | 3(stainless steel) |
Access port | 2pc. Ø 25 mm |
Casters | 4+(2 leveling feet) |
Data Logging/Interval/Memory capacity | USB/Record every 5 minutes / 10 years |
Backup battery | Yes |
Alarm | |
Temperature | High/Low temperature,High ambient temperature |
Electrical | Power failure , Low battery,Battery power detection failure |
System | Sensor failure,Communication error, Condenser overheating,Door ajar |
Accessory | |
Freezer rack model | ZKML304-552B-CT |
Freezer rack/box (quantity) | 16/400 |